Endobag separation - an ominous sign after endovascular aneurysm sealing with the Nellix device

Vasa. 2018 Jun;47(4):327-329. doi: 10.1024/0301-1526/a000699. Epub 2018 Mar 26.

Abstract

Our report aims to alert physicians performing endovascular aneurysm sealing with the Nellix endoprosthesis to the risk of rupture when endobag separation is noticed on computed tomography. Endobag separation when observed with acute presentation of abdominal pain is an indicator of imminent rupture and prompt interventional treatment should be undertaken.

Keywords: Endovascular aneurysm sealing; Nellix endoprosthesis; endobag; rupture.

Publication types

  • Case Reports
  • Review

MeSH terms

  • Abdominal Pain / etiology
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / surgery*
  • Aortic Rupture / diagnostic imaging
  • Aortic Rupture / etiology*
  • Aortography / methods
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Computed Tomography Angiography
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation*
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Failure*
  • Stents*
  • Treatment Outcome