Development of a Head Scanner for Proton CT

Nucl Instrum Methods Phys Res A. 2013 Jan 21:699:205-210. doi: 10.1016/j.nima.2012.04.029. Epub 2012 Apr 13.

Abstract

We describe a new head scanner developed for Proton Computed Tomography (pCT) in support of proton therapy treatment planning, aiming at reconstructing an accurate map of the stopping power (S.P.) in a phantom and, in the future, in patients. The system consists of two silicon telescopes which track the proton before and after the phantom/patient, and an energy detector which measures the residual energy or range of the proton to reconstruct the Water Equivalent Path Length (WEPL) in the phantom. Based on the experience of the existing prototype and extensive Geant4 simulations and CT reconstructions, the new pCT scanner will support clinically useful proton fluxes.