Cerebral aneurysms treated with flow-diverting stents: computational models with intravascular blood flow measurements

AJNR Am J Neuroradiol. 2014 Jan;35(1):143-8. doi: 10.3174/ajnr.A3624. Epub 2013 Jul 18.

Abstract

Background and purpose: Computational fluid dynamics modeling is useful in the study of the hemodynamic environment of cerebral aneurysms, but patient-specific measurements of boundary conditions, such as blood flow velocity and pressure, have not been previously applied to the study of flow-diverting stents. We integrated patient-specific intravascular blood flow velocity and pressure measurements into computational models of aneurysms before and after treatment with flow-diverting stents to determine stent effects on aneurysm hemodynamics.

Materials and methods: Blood flow velocity and pressure were measured in peri-aneurysmal locations by use of an intravascular dual-sensor pressure and Doppler velocity guidewire before and after flow-diverting stent treatment of 4 unruptured cerebral aneurysms. These measurements defined inflow and outflow boundary conditions for computational models. Intra-aneurysmal flow rates, wall shear stress, and wall shear stress gradient were calculated.

Results: Measurements of inflow velocity and outflow pressure were successful in all 4 patients. Computational models incorporating these measurements demonstrated significant reductions in intra-aneurysmal wall shear stress and wall shear stress gradient and a trend in reduced intra-aneurysmal blood flow.

Conclusions: Integration of intravascular dual-sensor guidewire measurements of blood flow velocity and blood pressure provided patient-specific computational models of cerebral aneurysms. Aneurysm treatment with flow-diverting stents reduces blood flow and hemodynamic shear stress in the aneurysm dome.

Publication types

  • Case Reports
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Blood Flow Velocity
  • Blood Pressure
  • Blood Vessel Prosthesis*
  • Cerebral Arteries / physiopathology*
  • Cerebral Arteries / surgery
  • Cerebrovascular Circulation*
  • Computer Simulation
  • Equipment Failure Analysis
  • Female
  • Humans
  • Intracranial Aneurysm / diagnosis
  • Intracranial Aneurysm / physiopathology*
  • Intracranial Aneurysm / surgery*
  • Male
  • Models, Cardiovascular*
  • Models, Neurological
  • Prosthesis Failure
  • Statistics as Topic
  • Stents*
  • Treatment Outcome