Adenosine triphosphate bioluminescence for bacteriologic surveillance and reprocessing strategies for minimizing risk of infection transmission by duodenoscopes

Gastrointest Endosc. 2017 Jun;85(6):1180-1187.e1. doi: 10.1016/j.gie.2016.10.035. Epub 2016 Nov 3.

Abstract

Background and aims: Recent outbreaks of duodenoscope-transmitted infections underscore the importance of adequate endoscope reprocessing. Adenosine triphosphate (ATP) bioluminescence testing allows rapid evaluation of endoscopes for bacteriologic/biologic residue. In this prospective study we evaluate the utility of ATP in bacteriologic surveillance and the effects of endoscopy staff education and dual cycles of cleaning and high-level disinfection (HLD) on endoscope reprocessing.

Methods: ATP bioluminescence was measured after precleaning, manual cleaning, and HLD on rinsates from suction-biopsy channels of all endoscopes and elevator channels of duodenoscopes/linear echoendoscopes after use. ATP bioluminescence was remeasured in duodenoscopes (1) after re-education and competency testing of endoscopy staff and subsequently (2) after 2 cycles of precleaning and manual cleaning and single cycle of HLD or (3) after 2 cycles of precleaning, manual cleaning, and HLD.

Results: The ideal ATP bioluminescence benchmark of <200 relative light units (RLUs) after manual cleaning was achieved from suction-biopsy channel rinsates of all endoscopes, but 9 of 10 duodenoscope elevator channel rinsates failed to meet this benchmark. Re-education reduced RLUs in duodenoscope elevator channel rinsates after precleaning (23,218.0 vs 1340.5 RLUs, P < .01) and HLD (177.0 vs 12.0 RLUs, P < .01). After 2 cycles of manual cleaning/HLD, duodenoscope elevator channel RLUs achieved levels similar to sterile water, with corresponding negative cultures.

Conclusions: ATP testing offers a rapid, inexpensive alternative for detection of endoscope microbial residue. Re-education of endoscopy staff and 2 cycles of cleaning and HLD decreased elevator channel RLUs to levels similar to sterile water and may therefore minimize the risk of transmission of infections by duodenoscopes.

MeSH terms

  • Adenosine Triphosphate / analysis*
  • Cross Infection / prevention & control*
  • Disease Transmission, Infectious / prevention & control*
  • Disinfection / methods*
  • Duodenoscopes / microbiology*
  • Endoscopy, Gastrointestinal
  • Equipment Contamination / prevention & control*
  • Equipment Reuse*
  • Humans
  • Luminescent Measurements*
  • Professional Competence