Background: Data regarding efficacy of various stent and embolic protection device (EPD) combinations to prevent stroke during carotid artery stenting (CAS) is limited. We compared post-procedure inpatient neurologic outcomes across various carotid stent-EPD platforms recorded in the Vascular Quality Initiative (VQI) registry.
Methods: We analyzed 13,786 consecutive CAS procedures in the VQI registry performed between January 1, 2005 and December 31, 2015. The most commonly used stent-EPD combinations (n = 5407) were included in the analysis. Post-procedure inpatient neurologic outcomes included (1) ipsilateral stroke/transient ischemic attack (TIA) and (2) any stroke/TIA. Multivariate generalized estimating equation regression analysis was performed, adjusting for age, sex, tobacco use, coronary artery disease, congestive heart failure, prior stroke/TIA, hypertension, history of carotid revascularization, and presence of a second ipsilateral stenosis >70%, to determine whether risk of outcomes differed according to device.
Results: Of 13,786 CAS procedures, Xact-Emboshield (n = 2,438, 17.6%), Precise-Angioguard (n = 1,480, 10.7%), Acculink-Accunet (n = 829, 6.01%), and Acculink-Emboshield (n = 660, 4.8%) were the most commonly used combinations, accounting for a total of 5,407 procedures. Inpatient event rates for ipsilateral stroke/TIA and any stroke/TIA were 1.9 and 2.7% in the Accunet-Acculink, 3.0 and 3.2% in Acculink-Emboshield, 3.2 and 4.1% in Precise-Angioguard and 2.2 and 3.0% in Xact-Emboshield. There was no evidence of difference in risk of ipsilateral stroke/TIA or any stroke/TIA across device combinations (P = 0.15 and P = 0.16, respectively).
Conclusion: CAS with current carotid stent-EPD combinations is associated with low rates of inpatient stroke/TIA. There is no statistically significant difference in rates of inpatient stroke/TIA across device combinations.
Keywords: carotid artery stenting; carotid stenosis; embolic protection; stroke.