Objective: Photoplethysmography (PPG) monitoring has been implemented in many portable and wearable devices we use daily for health and fitness tracking. Its simplicity and cost-effectiveness has enabled a variety of biomedical applications, such as continuous long-term monitoring of heart arrhythmias, fitness, and sleep tracking, and hydration monitoring. One major issue that can hinder PPG-based applications is movement artifacts, which can lead to false interpretations. In many implementations, noisy PPG signals are discarded. Misinterpreted or discarded PPG signals pose a problem in applications where the goal is to increase the yield of detecting physiological events, such as in the case of paroxysmal atrial fibrillation (AF)-a common episodic heart arrhythmia and a leading risk factor for stroke. In this work, we compared a traditional machine learning and deep learning approaches for PPG quality assessment in the presence of AF, in order to find the most robust method for PPG quality assessment.

Approach: The training data set was composed of 78 278 30 s long PPG recordings from 3764 patients using bedside patient monitors. Two different representations of PPG signals were employed-a time-series based (1D) one and an image-based (2D) one. Trained models were tested on an independent set of 2683 30 s PPG signals from 13 stroke patients.

Main results: ResNet18 showed a higher performance (0.985 accuracy, 0.979 specificity, and 0.988 sensitivity) than SVM and other deep learning approaches. 2D-based models were generally more accurate than 1D-based models.

Significance: 2D representation of PPG signal enhances the accuracy of PPG signal quality assessment.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Atrial Fibrillation / diagnosis*
  • Deep Learning*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Plethysmography*
  • Signal Processing, Computer-Assisted*
  • Support Vector Machine
  • Young Adult