Characterization of the PRAETORIAN score in Japanese patients undergoing subcutaneous implantable cardioverter-defibrillator implantation

J Cardiol. 2022 Nov;80(5):482-486. doi: 10.1016/j.jjcc.2022.06.014. Epub 2022 Jul 25.

Abstract

Background: The PRAETORIAN score was developed to evaluate the implant position and predict defibrillation success in patients implanted with a subcutaneous implantable cardioverter-defibrillator (S-ICD). However, usefulness of the PRAETORIAN score for Japanese patients is unknown.

Methods: We evaluated usefulness of this score, which was determined by width of sub-coil fat, sub-generator fat, and anterior positioning of the S-ICD generator by post-operative chest X-ray, in consecutive 100 Japanese S-ICD implanted patients [78 men, median age 59 (IQR 46.5-67.0) years, median body mass index (BMI) 24.2 (21.3-27.2) kg/m2].

Results: The median PRAETORIAN score was 30 (30-45) and 93 patients were classified as a low risk of conversion failure. The remaining seven were at an intermediate risk. Almost all patients were classified as an optimal pulse-generator position in the second and third steps of the PRAETORIAN score. The only difference observed was in the width of sub-coil fat in the first step. To further evaluate its significance, patients were divided into the Thicker group (sub-coil fat >1 coil width, n = 19) and the Thinner group (sub-coil fat ≤1 coil width, n = 81). BMI and post-shock impedance were both higher in the Thicker group than in the Thinner group [27.1 (25.6-31.6) versus 23.1 (20.9-25.7) kg/m2, p < 0.001, and 75 (68-88) versus 63 (55-74) Ω, p = 0.003, respectively]. During the median follow-up periods of 888 (523-1418) days, 7 patients experienced appropriate shock therapy for spontaneous ventricular tachyarrhythmias, who were all at a low risk. No conversion failure was observed. Inappropriate shock (IAS) occurred in 11 patients, and there was no difference in IAS rate between the Thicker group (n = 2) and the Thinner group (n = 9) (p = 0.747 by log-rank test).

Conclusions: Most Japanese patients were classified as at low risk of conversion failure. The PRAETORIAN score may be useful for the evaluation of conversion failure in Japanese S-ICD implanted patients.

Keywords: Body mass index; Japanese; Optimal implantation; PRAETORIAN score; Subcutaneous implantable cardioverter-defibrillator.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Body Mass Index
  • Defibrillators, Implantable* / adverse effects
  • Electric Countershock / adverse effects
  • Electric Impedance
  • Humans
  • Japan
  • Male
  • Middle Aged
  • Treatment Outcome