First-Attempt Intubation Success Among Emergency Medicine Trainees by Laryngoscopic Device and Training Year: A National Emergency Airway Registry Study

Ann Emerg Med. 2023 Jun;81(6):649-657. doi: 10.1016/j.annemergmed.2022.10.019. Epub 2023 Jan 18.

Abstract

Study objective: We compare intubation first-attempt success with the direct laryngoscope, hyperangulated video laryngoscope, and standard geometry video laryngoscope among emergency medicine residents at various postgraduate years (PGY) of training.

Methods: We analyzed prospective data from emergency department (ED) patients enrolled in the National Emergency Airway Registry from January 1, 2016 to December 31, 2018 using mixed-effects logistic regression to assess the association between PGY of training and first-attempt success by the device.

Results: Among 15,204 intubations performed by emergency medicine trainees, first-attempt success for PGY-1, PGY-2, and PGY3+ residents, respectively were: 78.8% (95% CI, 75.0 to 82.2%), 81.3% (79.4 to 83.0), and 83.6% (95% CI, 82.1 to 85.1) for direct laryngoscope; 87.2% (95% CI, 84.2 to 89.7), 90.4% (95% CI, 88.8 to 91.9%), and 91.2% (95% CI, 89.8 to 92.5%) for hyperangulated video laryngoscope; and 88.7% (95% CI, 86.1 to 90.9), 90.2% (95% CI, 88.7 to 91.5%), and 94.6% (95% CI 93.9 to 95.3%) for standard geometry video laryngoscope. Direct laryngoscope first-attempt success improved for PGY-2 (adjusted odds ratio [aOR],1.41; 95% CI, 1.09 to 1.82) and PGY-3+ (aOR, 1.76; 1.36 to 2.27) trainees compared to PGY-1. Hyperangulated video laryngoscope success also improved for PGY-2 (aOR, 1.51; 1.1 to 2.05) and PGY-3+ (aOR, 1.56; 1.15 to 2.13) trainees compared to PGY-1. For the standard geometry video laryngoscope, only PGY-3+ (aOR, 1.72; 1.25 to 2.36) was associated with improved first-attempt success compared to PGY-1.

Conclusion: Each laryngoscopy device class was associated with improvement in first-attempt success as training progressed. The video laryngoscope outperformed the direct laryngoscope for all operator groups, and PGY-1 trainees achieved higher first-attempt success using a standard geometry video laryngoscope than PGY-3+ trainees using a direct laryngoscope. These findings support the conjecture that in adult patients, a direct laryngoscope should not be routinely used for the first intubation attempt unless clinical circumstances, such as the presence of a soiled airway, would favor its success. These findings need to be validated with prospective randomized clinical trials.

MeSH terms

  • Adult
  • Emergency Medicine*
  • Emergency Service, Hospital
  • Humans
  • Intubation, Intratracheal
  • Laryngoscopes*
  • Laryngoscopy
  • Prospective Studies
  • Registries
  • Video Recording