The AccuFlow sensor: a novel digital health tool to assess intrapartum blood loss at cesarean delivery

J Perinat Med. 2023 May 9;51(8):997-1005. doi: 10.1515/jpm-2023-0101. Print 2023 Oct 26.

Abstract

Objectives: During obstetric hemorrhage, peripheral vasoconstriction maintains heart rate and blood pressure until compensatory mechanisms are overwhelmed and patients deteriorate rapidly. Real-time perfusion measurements could quantify vasoconstriction, improving early recognition of hemorrhage and facilitating early intervention to reduce morbidity and mortality. The AccuFlow device makes rapid, non-invasive, quantitative measurements of perfusion, but has not been studied for hemorrhage detection or used in surgical settings. This study evaluated feasibility, tolerability, and preliminary efficacy of the AccuFlow for assessment of blood loss at cesarean delivery (CD).

Methods: In this pilot study, sensors were applied to the wrist, forearm, bicep, and chest wall of 25 patients undergoing scheduled CD. Postoperatively, sensors were removed and patients rated the AccuFlow and the standard anesthesia monitoring equipment on a validated comfort rating scale for wearable computers (CRS). Blood loss was estimated by the surgical team (EBL) and calculated from change in hematocrit, weight, and height (CBL). CRS scores were compared via Wilcoxon signed ranks tests. Coefficients of correlation between sensor readings and CBL, and between EBL and CBL, were compared using Fisher's R-to-z transformation.

Results: There were no safety events; no participants requested device removal. CRS ratings of the AccuFlow and the standard monitoring equipment were similar (7.2 vs. 8.8, p=0.25). Change in wrist perfusion from delivery to dressing placement was more strongly correlated with CBL than was EBL (R=-0.48 vs. R=0.087, p=0.03).

Conclusions: The AccuFlow sensor is well-tolerated and shows promise in detecting intrapartum hemorrhage, though larger studies are needed.

Keywords: AccuFlow; intrapartum hemorrhage; obstetric hemorrhage; perfusion sensor; postpartum hemorrhage; surgical blood loss.