IMAGINER: improving accuracy with a mixed reality navigation system during placement of external ventricular drains. A feasibility study

Neurosurg Focus. 2024 Jan;56(1):E8. doi: 10.3171/2023.10.FOCUS23554.

Abstract

Objective: The placement of a ventricular catheter, that is, an external ventricular drain (EVD), is a common and essential neurosurgical procedure. In addition, it is one of the first procedures performed by inexperienced neurosurgeons. With or without surgical experience, the placement of an EVD according to anatomical landmarks only can be difficult, with the potential risk for inaccurate catheter placement. Repeated corrections can lead to avoidable complications. The use of mixed reality could be a helpful guide and improve the accuracy of drain placement, especially in patients with acute pathology leading to the displacement of anatomical structures. Using a human cadaveric model in this feasibility study, the authors aimed to evaluate the accuracy of EVD placement by comparing two techniques: mixed reality and freehand placement.

Methods: Twenty medical students performed the EVD placement procedure with a Cushing's ventricular cannula on the right and left sides of the ventricular system. The cannula was placed according to landmarks on one side and with the assistance of mixed reality (Microsoft HoloLens 2) on the other side. With mixed reality, a planned trajectory was displayed in the field of view that guides the placement of the cannula. Subsequently, the actual position of the cannula was assessed with the help of a CT scan with a 1-mm slice thickness. The bony structure as well as the left and right cannula positions were registered to the CT scan with the planned target point before the placement procedure. CloudCompare software was applied for registration and evaluation of accuracy.

Results: EVD placement using mixed reality was easily performed by all medical students. The predefined target point (inside the lateral ventricle) was reached with both techniques. However, the scattering radius of the target point reached through the use of mixed reality (12 mm) was reduced by more than 54% compared with the puncture without mixed reality (26 mm), which represents a doubling of the puncture accuracy.

Conclusions: This feasibility study specifically showed that the integration and use of mixed reality helps to achieve more than double the accuracy in the placement of ventricular catheters. Because of the easy availability of these new tools and their intuitive handling, we see great potential for mixed reality to improve accuracy.

Keywords: external ventricular drain; mixed reality; ventriculostomy.