Brain Arteriovenous Malformation In Vitro Model for Transvenous Embolization Using 3D Printing and Real Patient Data

AJNR Am J Neuroradiol. 2024 May 9;45(5):612-617. doi: 10.3174/ajnr.A8188.

Abstract

Background and purpose: Transvenous embolization has emerged as a novel technique for treating selected brain AVMs with high reported occlusion rates. However, it requires anatomic and technical skills to be successful and to ensure patient safety. Therefore, training and testing are essential for preparing clinicians to perform these procedures. Our aim was to develop and test a novel, patient-specific brain AVM in vitro model for transvenous embolization by using 3D printing technology.

Materials and methods: We developed a brain AVM in vitro model based on real patient data by using stereolithography resin 3D printing. We created a closed pulsed circuit with flow passing from the arterial side to the venous side, and we tested the effect of mean arterial pressure on retrograde nidal filling with contrast injections. Transvenous embolization simulations were conducted for each of the 12 identical models divided into 2 groups (2×6). This involved the use of an ethylene-vinyl alcohol liquid embolic agent injected through microcatheters either without or with a coil in the vein (groups 1 and 2, respectively).

Results: Retrograde contrast advance to nidus was directly related to lower mean arterial pressure. Transvenous embolization tests with a liquid embolic agent adequately reproduced the usual embolization plug and push technique. We found no differences between the 2 group conditions, and additional venous coil neither increased nidus penetration nor reduced injection time in the model (57.6 versus 61.2% nidus occlusion rate, respectively).

Conclusions: We were able to develop and test a functional in vitro brain AVM model for transvenous embolization by using 3D printing to emulate its conditions and characteristics. Better contrast penetration was achieved with less mean arterial pressure, and no embolization advantage was found by adding coil to the vein in this model.

MeSH terms

  • Embolization, Therapeutic* / methods
  • Humans
  • Intracranial Arteriovenous Malformations* / diagnostic imaging
  • Intracranial Arteriovenous Malformations* / therapy
  • Models, Anatomic
  • Printing, Three-Dimensional*