CONFINE-MAS: a magic-angle spinning NMR probe that confines the sample in case of a rotor explosion

J Biomol NMR. 2018 Dec;72(3-4):171-177. doi: 10.1007/s10858-018-0218-x. Epub 2018 Dec 10.

Abstract

Magic-angle spinning (MAS) is mandatory in solid-state NMR experiments to achieve resolved spectra. In rare cases, instabilities in the rotation or damage of either the rotor or the rotor cap can lead to a so called "rotor crash" involving a disintegration of the sample container and possibly the release of an aerosol or of dust. We present a modified design of a 3.2 mm probe with a confining chamber which in case of a rotor crash prevents the release of aerosols and possibly hazardous materials. 1D and 2D NMR experiments show that such a hazardous material-confining MAS probe ("CONFINE-MAS" probe) has a similar sensitivity compared to a standard probe and performs equally well in terms of spinning stability. We illustrate the CONFINE-MAS probe properties and performance by application to a fungal amyloid.

Keywords: Containment; Magic-angle spinning; Safety; Solid-state NMR.

MeSH terms

  • Amyloid
  • Containment of Biohazards / methods
  • Equipment Failure*
  • Fungal Proteins
  • Nuclear Magnetic Resonance, Biomolecular / instrumentation*
  • Safety*

Substances

  • Amyloid
  • Fungal Proteins